How to test/check Lab-Jack work condition , testing procedures of LabJack

Lab-jack testing procedure

Model no- U3 _HV

               

1.       Connect Cable laptop to Lab-jack

2.       Open software – LJ  control panel

 

              

 

 

3.       Select Model , click and find devices ( F3 )

4.       Then click on USB

5.       Lab-jack Model Number/ Serial Number  appear  on the NEW  window

6.       Cross check on Hardware ( Lab-jack )

7.       Check Default configuration




8.       Click on Test Tap.

 



 

9.        Pin details  check as per data sheets , and check functionality of each pin

10.   Using Multimeter , using different pin operation as per Data sheet

11.   If all pin work as per data sheet then ok

12.   Check supply on Lab-jack by multimeter

13.   Check output _ FIO Channel

14.   Select FIO4 , click Do

15.   Check on Hardware  FIO4-GND , should be 3.3 V , similar to other pin

16.   Check input , select input E.g._ AIN0, AIN1, AIN2, AIN3, connect to Ground , Voltage state should be Zero 0.0000 appear , similar to other pin

17.   Check 15 pin functionality, check voltage state according to data sheet with respect to pin number and GND , should be 3.3 V.

 

Thanks // Om prakash Gupta 